Remote control commands
R&S
®
RTO6
1450User Manual 1801.6687.02 ─ 05
Usage: Asynchronous command
REFCurve<m>:POSition <V
ertPosi>
Moves the reference waveform and its horizontal axis up or down in the diagram, if ver-
tical scaling is set to "Independent"
(see REFCurve<m>:VMODe on page 1449).
Suffix:
<m>
.
1..4
Reference waveform
Parameters:
<VertPosi> Range: -100E+24 to 100E+24
Increment: 0.02
*RST: 0
Default unit: div
Usage: Asynchronous command
REFCurve<m>:TOORignal
Restores the original vertical settings of the reference waveform (vertical scale, posi-
tion, and offset), if vertical scaling is set to independent (see REFCurve<m>:
VMODe
on page 1449).
Suffix:
<m>
.
1..4
Usage: Event
Asynchronous command
REFCurve<m>:RESCale:VERTical:STATe <State>
Enables and disables the vertical stretching. Stretching changes the display of the
waveform independent of the vertical scale and position.
Suffix:
<m>
.
1..4
Reference waveform
Parameters:
<State> ON | OFF
*RST: OFF
Usage: Asynchronous command
REFCurve<m>:RESCale:VERTical:FACTor <ScaleFactor>
Defines the vertical stretching factor. A factor greater than 1 stretches the waveform
vertically
, a factor lower than 1 compresses the curve.
Waveform analysis

Table of Contents