Remote control commands
R&S
®
RTO6
1451User Manual 1801.6687.02 ─ 05
Suffix:
<m>
.
1..4
Reference waveform
Parameters:
<ScaleFactor> Range: -1E+6 to 1E+6
Increment: 0.1
*RST: 1
Usage: Asynchronous command
REFCurve<m>:RESCale:VERTical:OFFSet <Offset>
Moves the reference waveform vertically
. Like vertical offset of channel waveforms, the
offset of a reference waveform is subtracted from the measured value.
Suffix:
<m>
.
1..4
Reference waveform
Parameters:
<Offset> Negative values shift the waveform up, positive values shift it
down.
Range: -100E+24 to 100E+24
Increment: 1E-6
*RST: 0
Default unit: V
Usage: Asynchronous command
REFCurve<m>:HMODe <HorizontalMode>
Selects the coupling of horizontal settings.
Suffix:
<m>
.
1..4
Reference waveform
Parameters:
<HorizontalMode> ORIGinal | COUPled
ORIGinal
Horizontal scaling and reference point of the source waveform
are used.
COUPled
The current horizontal settings of the diagram are used.
*RST: ORIGinal
Usage: Asynchronous command
Waveform analysis