Remote control commands
R&S
®
RTO6
1449User Manual 1801.6687.02 ─ 05
REFCurve<m>:RESCale:VERTical:OFFSet....................................................................
1451
REFCurve<m>:
HMODe................................................................................................1451
REFCurve<m>:RESCale:HORizontal:STATe...................................................................1452
REFCurve<m>:RESCale:HORizontal:FACTor................................................................. 1452
REFCurve<m>:RESCale:HORizontal:OFFSet.................................................................1452
REFCurve<m>:RESCale:HORizontal:POSition............................................................... 1453
REFCurve<m>:RESTore
Resets the time scale and the reference point to the original values of the reference
waveform.
Suffix:
<m>
.
1..4
Reference waveform
Usage: Event
Asynchronous command
REFCurve<m>:VMODe <VerticalMode>
Selects the coupling of vertical settings.
Suffix:
<m>
.
1..4
Reference waveform
Parameters:
<V
erticalMode> COUPled | INDependent
COUPled
V
ertical position and scale of the source are used.
INDependent
Scaling and position can be set specific to the reference wave-
form.
*RST
: INDependent
Usage: Asynchronous command
REFCurve<m>:SCALe <VerticalScale>
Sets the scale factor for the reference waveform if vertical scaling is set to
"Independ-
ent" (see REFCurve<m>:VMODe on page 1449).
Suffix:
<m>
.
1..4
Reference waveform
Parameters:
<VerticalScale> Range: 1E-15 to 100E+24
Increment: 10E-6
*RST: 0.5