Remote control commands
R&S
®
RTO6
1520User Manual 1801.6687.02 ─ 05
Usage: Asynchronous command
MEASurement<m>:OPTical:LCHeck<n>:LOWer:LIMit <LowerLimit>
MEASurement<m>:
OPT
ical:LCHeck<n>:UPPer:LIMit <UpperLimit>
Define the lower/ upper limit for limit checking of the optical measurements.
The range is defined with MEASurement<m>:OPTical:LCHeck<n>:VALid.
Suffix:
<m>
.
1..10
See "Selection of the measurement group: MEASurement<m>"
on page 1479.
<n> 1..2
Parameters:
<UpperLimit> Range: -100 to 100
Increment: 10E-6
*RST: 0
Usage: Asynchronous command
MEASurement<m>:OPTical:LCHeck<n>:LOWer:MARGin <LowerMargin>
MEASurement<m>:OPT
ical:LCHeck<n>:UPPer:MARGin <UpperMargin>
Define the lower/ upper margin for limit checking of the optical measurements.
The range is defined with MEASurement<m>:OPTical:LCHeck<n>:VALid.
Suffix:
<m>
.
1..10
See "Selection of the measurement group: MEASurement<m>"
on page 1479.
<n> 1..2
Parameters:
<UpperMargin> Range: -100 to 100
Increment: 10E-6
*RST: 0
Usage: Asynchronous command
MEASurement<m>:OPTical:LCHeck<n>:VALid <ValidRange>
Selects the condition for the limit check of the optical measurement.
Suffix:
<m>
.
1..10
See
"Selection of the measurement group: MEASurement<m>"
on page 1479.
Automatic measurements