EasyManua.ls Logo

R&S RTO6 - Page 1548

R&S RTO6
2677 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Remote control commands
R&S
®
RTO6
1548User Manual 1801.6687.02 ─ 05
MEASurement<m>:GATE:CURSor <Cursorset>
Selects the cursor set to be used for measurement gating. The gate area is defined by
the cursor lines.
Suffix:
<m>
.
1..10
See "Selection of the measurement group: MEASurement<m>"
on page 1479.
Parameters:
<Cursorset>
CURSOR1 | CURSor1 | CURSOR2 | CURSor2 | CURSOR3 |
CURSor3 | CURSOR4 | CURSor4
CURSOR1 = CURSor1, CURSOR2 = CURSor2, CURSOR3 =
CURSor3, CURSOR4 = CURSor4
*RST: CURSOR1
Usage: Asynchronous command
MEASurement<m>:GATE:CCOupling <CursorCoupling>
Enables the cursor coupling for automatic measurements.
Select the cursor set to be used with MEASurement<m>:GATE:CURSor.
Suffix:
<m>
.
1..10
See
"Selection of the measurement group: MEASurement<m>"
on page 1479.
Parameters:
<CursorCoupling>
ON | OFF
*RST: OFF
Usage: Asynchronous command
MEASurement<m>:GATE:ZCOupling <ZoomCoupling>
If enabled, the gate area is defined identically to the zoom area for the zoom diagram.
If enabled, define the zoom area to be used as gate with MEASurement<m>:GATE:
ZDIagram.
Suffix:
<m>
.
1..10
See
"Selection of the measurement group: MEASurement<m>"
on page 1479.
Parameters:
<ZoomCoupling>
ON | OFF
*RST: OFF
Usage: Asynchronous command
Automatic measurements

Table of Contents

Other manuals for R&S RTO6

Related product manuals