Remote control commands
R&S
®
RTO6
1679User Manual 1801.6687.02 ─ 05
TRIGger<m>:I2C:AMODe.............................................................................................
1680
TRIGger<m>:I2C:ACONdition
....................................................................................... 1681
TRIGger<m>:I2C:ADDRess.......................................................................................... 1681
TRIGger<m>:I2C:ADDTo.............................................................................................. 1681
TRIGger<m>:I2C:ADOR<n>:ENABle............................................................................. 1681
TRIGger<m>:I2C:ADOR<n>:ADRType...........................................................................1682
TRIGger<m>:I2C:ADOR<n>[:VALue]............................................................................. 1682
TRIGger<m>:I2C:ADOR<n>:RWBit................................................................................1682
TRIGger<m>:I2C:DPOPerator.......................................................................................1683
TRIGger<m>:I2C:DPOSition......................................................................................... 1683
TRIGger<m>:I2C:DPTO................................................................................................1683
TRIGger<m>:I2C:DCONdition....................................................................................... 1683
TRIGger<m>:I2C:DMIN................................................................................................ 1684
TRIGger<m>:I2C:DMAX............................................................................................... 1684
TRIGger<m>:I2C:TYPE <Type>
Selects the trigger type for I²C analysis.
Parameters:
<Type> STARt | REPStart | STOP | NACK | ADDRess | ADOR | ADAT
ST
ARt
Start condition
REPStart
Repeated start - the start condition occurs without previous stop
condition.
NACK
Missing acknowledge bit. T
o localize specific missing acknowl-
edge bits, use TRIGger<m>:I2C:ADNack, TRIGger<m>:
I2C:DWNack, and TRIGger<m>:I2C:DRNack.
ADDRess
Triggers on one specific address
ADOR
Triggers on an OR combination with up to four four address con-
ditions.
ADAT
Triggers on a combination of address and data condition.
STOP
Stop condition, end of frame
*RST: STARt
Usage: Asynchronous command
TRIGger<m>:I2C:ACCess <RWBitAddress>
Sets the trigger condition for the R/W bit - the transfer direction of the data.