EasyManua.ls Logo

R&S RTO6 - Page 1685

R&S RTO6
2677 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Remote control commands
R&S
®
RTO6
1685User Manual 1801.6687.02 ─ 05
BUS<m>:I2C:FRAMe<n>:BYTE<o>:COMPlete?.............................................................
1691
BUS<m>:I2C:FRAMe<n>:BYTE<o>:ST
ARt?...................................................................1692
BUS<m>:I2C:FRAMe<n>:BYTE<o>:VALue?...................................................................1692
BUS<m>:I2C:FRAMe<n>:DATA?
Returns the data words of the specified frame.
Suffix:
<m>
.
1..4
Selects the serial bus.
<n> *
Selects the frame.
Return values:
<Data> Comma-separated list of integer values (N, D1, D2,…, DN). N is
the number of bytes in the frame, and D1…DN are the values of
the bytes.
Example:
BUS:I2C:FRAMe4:DATA?
<-- 3,
74,164,18
Usage: Query only
BUS<m>:I2C:FCOunt?
Returns the number of decoded frames.
Suffix:
<m>
.
1..4
Selects the serial bus.
Return values:
<Count> Total number of decoded frames.
Usage: Query only
BUS<m>:I2C:FRAMe<n>:AACCess?
Returns the address acknowledge bit value for the indicated frame.
Suffix:
<m>
.
1..4
Selects the serial bus.
<n> *
Selects the frame.
Return values:
<AddressAckBit> INComplete | ACK | NACK | EITHer
*RST: INComplete
Usage: Query only
Protocols

Table of Contents

Other manuals for R&S RTO6

Related product manuals