Remote control commands
R&S
®
RTO6
1740User Manual 1801.6687.02 ─ 05
BUS<m>:I2S:DATA:POLarity <SDataPolarity>
Defines the interpretation of high and low signal states on the data line.
Suffix:
<m>
.
1..4
Selects the serial bus.
Parameters:
<SDataPolarity> ACTLow | ACTHigh
ACTHigh: HIGH = 1 and LOW = 0
ACTLow: HIGH = 0 and LOW = 1
*RST: ACTHigh
BUS<m>:I2S:TCOupling <Coupling>
Sets all thresholds to the value of the clock threshold BUS<m>:I2S:CLOCk:
THReshold.
Suffix:
<m>
.
1..4
Selects the serial bus.
Parameters:
<Coupling>
ON | OFF
*RST: ON
BUS<m>:I2S:CLOCk:THReshold <SCLKThreshold>
Sets the threshold value for the clock line SCLK.
If BUS<m>:I2S:
TCOupling is ON, the command sets the threshold for all lines.
Suffix:
<m>
.
1..4
Selects the serial bus.
Parameters:
<SCLKThreshold> Range: -10 to 10
Increment: 1E-3
*RST: 1.6
Default unit: V
BUS<m>:I2S:WSELect:THReshold <WSThreshold>
Sets the threshold value for the word select and FSYNC lines.
Suffix:
<m>
.
1..4
Selects the serial bus.