Remote control commands
R&S
®
RTO6
1748User Manual 1801.6687.02 ─ 05
<n> *
Selects the frame.
Return values:
<FrameState> ERRor | OK | INSufficient
OK: the frame is valid.
ERRor: an error occured in the frame.
INSufficient: The frame is not completely contained in the acqui-
sition. The acquired part of the frame is valid.
Usage: Query only
BUS<m>:I2S:FRAMe<n>:
STARt?
BUS<m>:I2S:FRAMe<n>:
ST
OP?
Return the start time and stop time of the selected frame.
Suffix:
<m>
.
1..4
Selects the serial bus.
<n> *
Selects the frame.
Return values:
<Start>, <Stop> Range: -100E+24 to 100E+24
*RST: 0
Usage: Query only
BUS<m>:I2S:FRAMe<n>:LEFT:VALue?
BUS<m>:I2S:FRAMe<n>:RIGHt:V
ALue?
Return the data values of the left and right channel, respectively.
Suffix:
<m>
.
1..4
Selects the serial bus.
<n> *
Selects the frame.
Return values:
<Value> Comma-separated list of values. To set the value format, use
FORMat:
BPATtern
.
Range: 0 to 4294967295
*RST: 0
Usage: Query only
Protocols