Remote control commands
R&S
®
RTO6
1841User Manual 1801.6687.02 ─ 05
TRIGger<m>:LIN:IMIN <IdPattern>
Specifies a slave identifier pattern, or sets the the start value of an identifier range.
Parameters:
<IdPattern> Numeric or string pattern, see Chapter 24.4.6, "Bit pattern
parameter", on page 1286.
TRIGger<m>:LIN:IMAX <IdPatternTo>
Sets the the end value of an identifier range if TRIGger<m>:
LIN:ICONdition is set
to INRange or OORange.
Parameters:
<IdPatternTo> Numeric or string pattern, see Chapter 24.4.6, "Bit pattern
parameter", on page 1286.
TRIGger<m>:LIN:DCONdition <DataOperator>
Sets the operator to set a specific data pattern or a data pattern range.
Parameters:
<DataOperator> EQUal | NEQual | LTHan | LETHan | GTHan | GETHan |
INRange | OORange
EQUal | NEQual | LTHan | LETHan | GTHan | GETHan
Equal, Not equal, Less than, Less or equal than, Greater Than,
Greater or equal than. These conditions require one data pattern
to be set with
TRIGger<m>:LIN:DMIN.
INRange | OORange
In range / Out of range: Set the minimum and maximum value of
the range with TRIGger<m>:LIN:DMIN and TRIGger<m>:
LIN:DMAX
*RST: EQUal
TRIGger<m>:LIN:DMIN <DataPattern>
Specifies a data pattern, or sets the the start value of a data pattern range.
Parameters:
<DataPattern> Numeric or string pattern, see Chapter 24.4.6, "Bit pattern
parameter", on page 1286. The string parameter accepts the bit
value X (don't care).
TRIGger<m>:LIN:DMAX <DataPatternTo>
Sets the the end value of an data range if TRIGger<m>:
LIN:DCONdition is set to
INRange or OORange.