Remote control commands
R&S
®
RTO6
1840User Manual 1801.6687.02 ─ 05
SYNC
Start of the frame, triggers on the stop bit of the sync field.
ID
Sets the trigger to one specific identifier or an identifier range.
T
o set the identifier
, use TRIGger<m>:LIN:ICONdition,
TRIGger<m>:LIN:IMIN on page 1841, and TRIGger<m>:
LIN:IMAX on page 1841.
IDOR
Triggers on an OR combination with up to four identifier condi-
tions. For each identifier condition, enable it with TRIGger<m>:
LIN:IDOR<n>:ENABle and set the value with TRIGger<m>:
LIN:IDOR<n>[:VALue]
IDDT
Combination of identifier and data conditions
To set the identifier condition, use TRIGger<m>:LIN:
ICONdition, TRIGger<m>:LIN:IMIN, and TRIGger<m>:
LIN:IMAX.
To set the data condition, use TRIGger<m>:LIN:DCONdition,
TRIGger<m>:LIN:DMIN, and TRIGger<m>:LIN:DMAX.
WKFR
Wakeup frame
ERRC
Error condition. Define the error types with TRIGger<m>:LIN:
CHKSerror on page 1844, TRIGger<m>:LIN:IPERror, and
TRIGger<m>:LIN:SYERror
*RST: SYNC
TRIGger<m>:LIN:ICONdition <IdOperator>
Sets the operator to set a specific identifier or an identifier range.
Parameters:
<IdOperator> EQUal | NEQual | LTHan | LETHan | GTHan | GETHan |
INRange | OORange
EQUal | NEQual | L
THan | LETHan | GTHan | GETHan
Equal, Not equal, Less than, Less or equal than, Greater Than,
Greater or equal than. These condition require one identifier pat-
tern to be set with
TRIGger<m>:LIN:IMIN
INRange | OORange
In range / Out of range: Set the minimum and maximum value of
the range with TRIGger<m>:LIN:IMIN and TRIGger<m>:
LIN:IMAX
*RST: EQUal
Protocols