Remote control commands
R&S
®
RTO6
1856User Manual 1801.6687.02 ─ 05
Suffix:
<m>
.
1..4
Index of the identifier in an "identifier OR" condition
Parameters:
<IdPattern> Numeric or string pattern, see Chapter 24.4.6, "Bit pattern
parameter", on page 1286.
Parameters for setting and query:
<SearchName>
SEARch:TRIGger:LIN:DCONdition <SearchName>,<DataOperator>
SEARch:TRIGger:LIN:DCONdition? <SearchName>
Sets the operator to set a specific data pattern or a data pattern range.
Parameters:
<DataOperator>
EQUal | NEQual | LTHan | LETHan | GTHan | GETHan |
INRange | OORange
EQUal | NEQual | L
THan | LETHan | GTHan | GETHan
Equal, Not equal, Less than, Less or equal than, Greater Than,
Greater or equal than. These conditions require one data pattern
to be set with
SEARch:TRIGger:LIN:DMIN.
INRange | OORange
In range / Out of range: Set the minimum and maximum value of
the range with SEARch:TRIGger:LIN:DMIN and SEARch:
TRIGger:LIN:DMAX.
*RST: EQUal
Parameters for setting and query:
<SearchName>
SEARch:TRIGger:LIN:DMIN <SearchName>,<DataPattern>
SEARch:TRIGger:LIN:DMIN? <SearchName>
Specifies a data pattern, or sets the the start value of a data pattern range.
Parameters:
<DataPattern> Numeric or string pattern, see Chapter 24.4.6, "Bit pattern
parameter", on page 1286.
Parameters for setting and query:
<SearchName>
SEARch:TRIGger:LIN:DMAX <SearchName>,<DataPatternTo>
SEARch:TRIGger:LIN:DMAX?
<SearchName>
Sets the the end value of an identifier range if SEARch:TRIGger:LIN:DCONdition
is set to INRange or OORange.