Remote control commands
R&S
®
RTO6
1867User Manual 1801.6687.02 ─ 05
*RST:  SOURce1: C1W1, SOURce2: C2W1
BUS<m>:FLXRay:THReshold<n> <THResholds>
Sets the thresholds for the bus signal if the source type is SINGle or DIFFerential.
For LOGic source type, use BUS<m>:
FLXRay:THData 
on page 1867 and BUS<m>:
FLXRay:THENable on page 1867.
Suffix: 
<m>
.
1..4
Selects the serial bus.
<n> 1 | 2
1 = threshold high, 2 = threshold low
Parameters:
<THResholds> Differential or absolute voltage level, depending on the source
type. See BUS<m>:FLXRay:SRCType 
on page 1865.
BUS<m>:FLXRay:THENable <ThresholdEnable>
Sets the threshold for the enable line if the source type is LOGic.
Suffix: 
<m>
.
1..4
Selects the serial bus.
Parameters:
<ThresholdEnable> Range:  -12  to  12
Increment:  0.1
*RST:  2.65
Default unit: V
BUS<m>:FLXRay:THData <ThresholdData>
Sets the threshold for the data line if the source type is LOGic.
Suffix: 
<m>
.
1..4
Selects the serial bus.
Parameters:
<ThresholdData> Range:  -12  to  12
Increment:  0.1
*RST:  2.35
Default unit: V
BUS<m>:FLXRay:PRSingle <PresetSingleEnd>
Selects a default threshold voltage if BUS<m>:FLXRay:
SRCType is set to SINGle.
Protocols