Remote control commands
R&S
®
RTO6
1875User Manual 1801.6687.02 ─ 05
Parameters:
<CycleCountMin> Range: 0 to 63
Increment: 1
*RST: 0
TRIGger<m>:FLXRay:CMAX <CycleCountMax>
Sets the the end value of a cycle count range if the condition TRIGger<m>:FLXRay:
CENable
on page 1874 is set to INRange or OORange.
Parameters:
<CycleCountMax> Range: 0 to 63
Increment: 1
*RST: 0
TRIGger<m>:FLXRay:CSTep <CycleCountStep>
Specifies a step to trigger on each n-th cycle inside the given range. This allows for
specific triggering if slot multiplexing is used.
The condition TRIGger<m>:FLXRay:
CENable on page 1874 must be set to
INRange or OORange.
Parameters:
<CycleCountStep> Range: 1 to 63
Increment: 1
*RST: 1
TRIGger<m>:FLXRay:DPOPerator <DataPosOperator>
Sets the operator for the data position. You can defined an exact position, or a position
range.
Parameters:
<DataPosOperator>
ANY | OFF | EQUal | GETHan | INRange | RANGe
OFF = ANY
The data position is not relevant for the trigger condition.
EQUal | GETHan
Equal, Greater or equal than. These conditions require one data
position to be set with TRIGger<m>:FLXRay:DPOSition.
INRange = RANGe
In range: Set the minimum and maximum value of the range with
TRIGger<m>:
FLXRay:DPOSition and TRIGger<m>:
FLXRay:DPTO.
*RST: EQUal
TRIGger<m>:FLXRay:DPOSition <DataPosition>
Sets the number of data bytes to be skipped after start of the payload segment
Protocols