Remote control commands
R&S
®
RTO6
1876User Manual 1801.6687.02 ─ 05
Parameters:
<DataPosition> Range: 0 to 255
Increment: 1
*RST: 0
TRIGger<m>:FLXRay:DPTO <DataPositionTo>
Defines the last byte of interest, if the position operator TRIGger<m>:
FLXRay:
DPOPerator defines a range.
Parameters:
<DataPositionTo> Range: 0 to 255
Increment: 1
*RST: 0
TRIGger<m>:FLXRay:DCONdition <DataOperator>
Sets the operator to set a specific data pattern or a data pattern range.
Parameters:
<DataOperator> OFF | ANY | EQUal | NEQual | LTHan | LETHan | GTHan |
GETHan | INRange | OORange
OFF = ANY
The data position is not relevant for the trigger condition.
EQUal | NEQual | LTHan | LETHan | GTHan | GETHan
Equal, Not equal, Less than, Less or equal than, Greater Than,
Greater or equal than. These conditions require one data posi-
tion to be set with TRIGger<m>:FLXRay:DMIN.
INRange | OORange
In range, Out of range: Set the minimum and maximum value of
the range with TRIGger<m>:FLXRay:DMIN and TRIGger<m>:
FLXRay:DMAX.
*RST: EQUal
TRIGger<m>:FLXRay:DMIN <DataPattern>
Specifies a data pattern, or sets the the start value of a data pattern range.
Parameters:
<DataPattern> Numeric or string pattern, see Chapter 24.4.6, "Bit pattern
parameter", on page
1286. The string parameter accepts the bit
value X (don't care).
TRIGger<m>:FLXRay:DMAX <DataPatternTo>
Sets the the end value of an data range if the operator TRIGger<m>:
FLXRay:
DCONdition is set to INRange or OORange.
Protocols