EasyManua.ls Logo

R&S RTO6 - Page 1876

R&S RTO6
2677 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Remote control commands
R&S
®
RTO6
1876User Manual 1801.6687.02 ─ 05
Parameters:
<DataPosition> Range: 0 to 255
Increment: 1
*RST: 0
TRIGger<m>:FLXRay:DPTO <DataPositionTo>
Defines the last byte of interest, if the position operator TRIGger<m>:
FLXRay:
DPOPerator defines a range.
Parameters:
<DataPositionTo> Range: 0 to 255
Increment: 1
*RST: 0
TRIGger<m>:FLXRay:DCONdition <DataOperator>
Sets the operator to set a specific data pattern or a data pattern range.
Parameters:
<DataOperator> OFF | ANY | EQUal | NEQual | LTHan | LETHan | GTHan |
GETHan | INRange | OORange
OFF = ANY
The data position is not relevant for the trigger condition.
EQUal | NEQual | LTHan | LETHan | GTHan | GETHan
Equal, Not equal, Less than, Less or equal than, Greater Than,
Greater or equal than. These conditions require one data posi-
tion to be set with TRIGger<m>:FLXRay:DMIN.
INRange | OORange
In range, Out of range: Set the minimum and maximum value of
the range with TRIGger<m>:FLXRay:DMIN and TRIGger<m>:
FLXRay:DMAX.
*RST: EQUal
TRIGger<m>:FLXRay:DMIN <DataPattern>
Specifies a data pattern, or sets the the start value of a data pattern range.
Parameters:
<DataPattern> Numeric or string pattern, see Chapter 24.4.6, "Bit pattern
parameter", on page
1286. The string parameter accepts the bit
value X (don't care).
TRIGger<m>:FLXRay:DMAX <DataPatternTo>
Sets the the end value of an data range if the operator TRIGger<m>:
FLXRay:
DCONdition is set to INRange or OORange.