Remote control commands
R&S
®
RTO6
1880User Manual 1801.6687.02 ─ 05
Return values:
<FrameStop> Time
Range: -100E+24 to 100E+24
*RST: 0
Default unit: s
Usage: Query only
BUS<m>:FLXRay:FRAMe<n>:SYMBol?
Returns the symbolic label of the specified frame if the label list is enabled.
Suffix:
<m>
.
1..4
Selects the serial bus.
<n> *
Selects the number of the frame in the current acquisition, 1...n.
Return values:
<Label> String with symbolic name of the identifier
Example:
BUS:FLXRay:FRAMe2:SYMBol?
Response: Temperature
Usage: Query only
BUS<m>:FLXRay:FRAMe<n>:TYPE?
Returns the frame type of the selected frame.
Suffix:
<m>
.
1..4
Selects the serial bus.
<n> *
Selects the frame.
Return values:
<FrameType> UNKNown | STATic | DYNamic | WAKE | SYMBol
STA
Tic: frame of the static segment
DYNamic: frame of the dynamic segment
WAKE: frame contains wakeup pattern
SYMBol: frame contains a MTS or CAS symbol
*RST: STATic
Usage: Query only
BUS<m>:FLXRay:FRAMe<n>:DATA?