Remote control commands
R&S
®
RTO6
1881User Manual 1801.6687.02 ─ 05
Suffix:
<m>
.
1..4
Selects the serial bus.
<n> *
Selects the frame.
Return values:
<Data> Comma-separated list of integer values (N, D1, D2,…, DN). N is
the number of bytes in the frame, and D1…DN are the values of
the bytes.
Example:
BUS:FLXRay:FRAMe4:DATA?
<-- 4,
17,85,170,
85
Usage: Query only
BUS<m>:FLXRay:FRAMe<n>:FLAGs?
Returns the value of the indicator bits at the beginning of the header segment. The five
bits are read as one word.
Suffix:
<m>
.
1..4
Selects the serial bus.
<n> *
Selects the frame.
Return values:
<Flags> Range: 0 to 31
*RST: 0
Usage: Query only
BUS<m>:FLXRay:FRAMe<n>:ADID?
Returns the frame identifier
, the number of the slot in which the frame is transmitted.
Suffix:
<m>
.
1..4
Selects the serial bus.
<n> *
Selects the frame.
Return values:
<AddressID>
To set the value format, use FORMat:
BPATtern
on page 1294.
The values below range, increment and default are decimal
values.
Range: 0 to 2047
*RST: 0
Usage: Query only
Protocols