Remote control commands
R&S
®
RTO6
1950User Manual 1801.6687.02 ─ 05
Parameters:
<ErrorCRC> ON | OFF
*RST: ON
Parameters for setting and query:
<SearchName>
SEARch:TRIGger:CXPI:ERRor:DLC <SearchName>,<ErrorDLC>
SEARch:TRIGger:CXPI:ERRor:DLC? <SearchName>
Searches for a data length code error
, if SEARch:TRIGger:CXPI:TYPE is set to
ERRor.
Parameters:
<ErrorDLC> ON | OFF
*RST: ON
Parameters for setting and query:
<SearchName>
SEARch:TRIGger:CXPI:ERRor:IBS <SearchName>,<ErrorIBS>
SEARch:TRIGger:CXPI:ERRor:IBS? <SearchName>
Searches for an inter-byte-space error, if SEARch:TRIGger:
CXPI:TYPE is set to
ERRor.
Parameters:
<ErrorIBS> ON | OFF
*RST: ON
Parameters for setting and query:
<SearchName>
SEARch:TRIGger:CXPI:ERRor:IFS <SearchName>,<ErrorIFS>
SEARch:TRIGger:CXPI:ERRor:IFS? <SearchName>
Searches for an inter-frame-space error, if SEARch:TRIGger:CXPI:TYPE is set to
ERRor.
Parameters:
<ErrorIFS> ON | OFF
*RST: ON
Parameters for setting and query:
<SearchName>
SEARch:TRIGger:CXPI:ERRor:LENGth <SearchName>,<ErrorLength>
SEARch:TRIGger:CXPI:ERRor:LENGth? <SearchName>
Searches for a length error, if SEARch:TRIGger:CXPI:TYPE is set to ERRor.