Remote control commands
R&S
®
RTO6
1965User Manual 1801.6687.02 ─ 05
TRIGger<m>:MILStd:CDST:IMAX <InfoPatternMax>
Sets the end value of the info range if TRIGger<m>:
MILStd:CDST:
ICONdition is
set to INRange.
Parameters:
<InfoPatternMax> Numeric or string pattern, see Chapter 24.4.6, "Bit pattern
parameter", on page 1286. The string parameter accepts the bit
value X (don't care).
TRIGger<m>:MILStd:CDST:IMIN <InfoPatternMin>
Specifies an info or sets the the start value of an info range.
Parameters:
<InfoPatternMin> Numeric or string pattern, see Chapter 24.4.6, "Bit pattern
parameter", on page
1286. The string parameter accepts the bit
value X (don't care).
TRIGger<m>:MILStd:DATA:DCONdition <DataOperator>
Sets the operator to set a specific data pattern.
Parameters:
<DataOperator> EQUal | NEQual | LTHan | LETHan | GTHan | GETHan |
INRange | OORange
EQUal | NEQual | LTHan | LETHan | GTHan | GETHan
Equal, Not equal, Less than, Less or equal than, Greater Than,
Greater or equal than. These conditions require one pattern to
be set with
TRIGger<m>:MILStd:DATA:DMIN.
INRange | OORange
In range/Out of range: set the minimum and maximum value of
the range with TRIGger<m>:MILStd:DATA:DMIN and
TRIGger<m>:MILStd:DATA:DMAX.
*RST: EQUal
TRIGger<m>:MILStd:DATA:DMAX <DataPatternMax>
Sets the end value of a data pattern range if TRIGger<m>:MILStd:
DATA:
DCONdition is set to INRange or OORange.
Parameters:
<DataPatternMax> Numeric or string pattern, see Chapter 24.4.6, "Bit pattern
parameter", on page 1286. The string parameter accepts the bit
value X (don't care).
TRIGger<m>:MILStd:DATA:DMIN <DataPatternMin>
Specifies a data pattern or sets the the start value of a data pattern range.
Protocols