Remote control commands
R&S
®
RTO6
2153User Manual 1801.6687.02 ─ 05
SPECial
Sets the trigger to one out of four dif
ferent Special PID trigger
types: PREamble, ERR, SPLIT
, or PING.
ERRCond
Sets the trigger to one out of seven different error condition trig-
ger types: PID error, CRC5 error, CRC16 error, Bitstuffing error,
Unexpected PID error, SE1 error, or Glitching error.
*RST: SOP
TRIGger<m>:USB:ACONdition <OperatorAddress>
Sets the operator to set a specific address or an address range. The address values
are set with TRIGger<m>:
USB:
AMIN and TRIGger<m>:USB:AMAX.
Parameters:
<OperatorAddress> OFF | ANY | EQUal | NEQual | LTHan | LETHan | GTHan |
GETHan | INRange | OORange
EQUal | NEQual | LTHan | LETHan | GTHan | GETHan
Equal, Not equal, Less than, Less than or equal, Greater than,
Greater than or equal. These conditions require one address
value to be set using TRIGger<m>:USB:AMIN.
INRange | OORange
In range, out of range. These conditions require an address
range to be set using TRIGger<m>:USB:AMIN and
TRIGger<m>:USB:AMAX.
*RST: OFF
TRIGger<m>:USB:AMIN <Address>
Specifies the address, or sets the the start value of an address range. The string
parameter does not accept the bit value X (don't care).
Parameters:
<Address>
TRIGger<m>:USB:AMAX <AddressTo>
Sets the the end value of an address range if TRIGger<m>:
USB:ACONdition is set
to INRange or OORange. The string parameter does not accept the bit value X (don't
care).
Parameters:
<AddressTo>