Remote control commands
R&S
®
RTO6
2154User Manual 1801.6687.02 ─ 05
TRIGger<m>:USB:DATA <USBDataT
ype>
Sets the trigger to one out of four dif
ferent payload data types: DATA0, DATA1, DATA2,
or MDATA.
Parameters:
<USBDataType> D0 | D1 | D2 | MD
D0
Sets the trigger to the DAT
A0 field (even PID).
D1
Sets the trigger to the DATA1 field (odd PID).
D2
Sets the trigger to the DATA2 field (data packet PID for high-
speed, high bandwidth isochronous transaction in a microframe).
MD
Sets the trigger to the MDATA field (high-speed data packet PID
for split and high bandwidth isochronous transactions).
*RST: D0
TRIGger<m>:USB:DCONdition <OperatorData>
Sets the operator (equal or unequal) to set a specific payload data pattern.
Parameters:
<OperatorData> OFF | ANY | EQUal | NEQual
*RST: EQUal
TRIGger<m>:USB:DPOPerator <DataPosOperator>
Sets the operator (any or equal) for the payload data index position.
Parameters:
<DataPosOperator> ANY | OFF | EQUal
ANY = OFF
The position of the bit pattern within the payload data is not rele-
vant.
EQUal
Sets the operator for specifying a special start position for the bit
pattern within the payload data.
*RST: ANY
TRIGger<m>:USB:DPOSition <DataPosition>
Specifies the position within a payload data packet, in which a special data pattern is to
be searched.