Remote control commands
R&S
®
RTO6
2155User Manual 1801.6687.02 ─ 05
Parameters:
<DataPosition> Range: 0 to 1024
Increment: 1
*RST: 0
TRIGger<m>:USB:ECONdition <EndpointOptor>
Sets the operator to set a specific endpoint or an endpoint range. The endpoint values
are set with TRIGger<m>:USB:
EMIN and TRIGger<m>:USB:EMAX.
Parameters:
<EndpointOptor> OFF | ANY | EQUal | NEQual | LTHan | LETHan | GTHan |
GETHan | INRange | OORange
EQUal | NEQual | LTHan | LETHan | GTHan | GETHan
Equal, Not equal, Less than, Less than or equal, Greater than,
Greater than or equal. These conditions require one endpoint
value to be set using TRIGger<m>:USB:EMIN.
INRange | OORange
In range, out of range. These conditions require a range of end-
point values to be set using TRIGger<m>:USB:EMIN and
TRIGger<m>:USB:EMAX.
*RST: OFF
TRIGger<m>:USB:EMIN <Endp>
Specifies the endpoint, or sets the the start value of an endpoint range.
Parameters:
<Endp>
TRIGger<m>:USB:EMAX <EndpTo>
Sets the the end value of an endpoint range if TRIGger<m>:
USB:ECONdition is set
to INRange or OORange.
Parameters:
<EndpTo>
TRIGger<m>:USB:ERRC <ErrorType>
Sets the trigger to an error condition type.
Parameters:
<ErrorType> ANY | PIDerror | CRC5error | CRC16error | BTST | UNEXpid |
SE1error | GLITcherr
ANY
T
riggers on any of the errors listed below.