Remote control commands
R&S
®
RTO6
2171User Manual 1801.6687.02 ─ 05
Parameters for setting and query:
<SearchName>
SEARch:TRIGger:USB:DATA <SearchName>,<USBDataT
ype>
SEARch:TRIGger:USB:DA
TA? <SearchName>
Defines, which data packet type is searched for: "DATA0", "DATA1", "DATA2", or
"MDATA", as well as "Any" data packet.
Parameters:
<USBDataType> ANY | D0 | D1 | D2 | MD
ANY
Searches for any of the data packet types listed below
D0
Searches for a DAT
A0 packet (even PID)
D1
Searches for a DATA1 packet (odd PID)
D2
Searches for a DATA2 packet (high-speed data packet for high
bandwidth isochronous transaction in a microframe)
MD
Searches for an MDATA packet (high-speed data packet for split
and high bandwidth isochronous transactions)
*RST: D0
Parameters for setting and query:
<SearchName>
SEARch:TRIGger:USB:DCONdition <SearchName>,<OperatorData>
SEARch:TRIGger:USB:DCONdition? <SearchName>
Sets the operator (equal or unequal) to set a specific payload data pattern.
Parameters:
<OperatorData> OFF | ANY | EQUal | NEQual
*RST: EQUal
Parameters for setting and query:
<SearchName>
SEARch:TRIGger:USB:DPOPerator <SearchName>,<DataPosOperator>
SEARch:TRIGger:USB:DPOPerator?
<SearchName>
Sets the operator (any or equal) for the payload data index position.