EasyManua.ls Logo

R&S RTO6 - Page 2175

R&S RTO6
2677 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Remote control commands
R&S
®
RTO6
2175User Manual 1801.6687.02 ─ 05
SEARch:TRIGger:USB:PCONdition <SearchName>,<OperatorPort>
SEARch:TRIGger:USB:PCONdition? <SearchName>
Sets the operator to set a specific port number or a port number range. The port num-
ber values are set with SEARch:TRIGger:
USB:
PMIN and SEARch:TRIGger:USB:
PMAX.
Parameters:
<OperatorPort> OFF | ANY | EQUal | NEQual | LTHan | LETHan | GTHan |
GETHan | INRange | OORange
EQUal | NEQual | LTHan | LETHan | GTHan | GETHan
Equal, Not equal, Less than, Less than or equal, Greater than,
Greater than or equal. These conditions require one port number
to be set using
SEARch:TRIGger:USB:PMIN.
INRange | OORange
In range, out of range. These conditions require a port number
range to be set using SEARch:TRIGger:USB:PMIN and
SEARch:TRIGger:USB:PMAX.
*RST: OFF
Parameters for setting and query:
<SearchName>
SEARch:TRIGger:USB:PMIN <SearchName>,<Port>
SEARch:TRIGger:USB:PMIN? <SearchName>
Specifies a port number
, or sets the start value of a port number range.
Parameters:
<Port>
Parameters for setting and query:
<SearchName>
SEARch:TRIGger:USB:PMAX <SearchName>,<PortTo>
SEARch:TRIGger:USB:PMAX? <SearchName>
Sets the the end value of a port number range if TRIGger<m>:USB:PCONdition is
set to INRange or OORange.
Parameters:
<PortTo>
Parameters for setting and query:
<SearchName>
SEARch:TRIGger:USB:PIDerror <SearchName>,<PIDError>
SEARch:TRIGger:USB:PIDerror?
<SearchName>
Defines, whether a search for any PID error shall be activated or not.
Protocols

Table of Contents

Other manuals for R&S RTO6

Related product manuals