EasyManua.ls Logo

R&S RTO6 - Page 225

R&S RTO6
2677 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Triggers
R&S
®
RTO6
225User Manual 1801.6687.02 ─ 05
Actions on trigger
The trigger can initiate several actions, each time a trigger occurs. To activate an
action, set it to "On trigger". The following actions are available:
"Beep"
Generates a beep sound.
"Save screen-
shot"
Saves a screenshot according to settings in
"Menu" > "Save/Recall" >
"Save" tab > "Screenshot".
"Save Wave-
form"
Saves the waveform data to a file according to settings in "Menu" >
"Save/Recall" > "Save" tab >
"Waveforms".
"Trigger out"
Selects, if a pulse is provided to the [Trigger Out] connector on the
rear panel.
Adjust the pulse with "Set trigger out", see
"Trigger out signal setup"
on page 225.
"Report"
Creates and saves a report using the settings defined in "Menu" key
> "Save/Recall" key >
"Save" tab > "Report".
"Start executa-
ble"
Starts an external application. Tap "Set executable" to set the applica-
tion path and parameters.
See: Chapter 4.8.3, "External application", on page 107.
Remote command:
TRIGger<m>:EVENt:
BEEP on page 1431
TRIGger<m>:EVENt:PRINt on page 1432
TRIGger<m>:EVENt:WFMSave on page 1432
TRIGger<m>:EVENt:RUNexec on page 1432
TRIGger<m>:OUT:ACTion on page 1433
Trigger out signal setup
Defines the pulse that is provided to the
[T
rigger Out] connector on the rear panel.
A trigger out pulse can be provided either when a trigger occurs, or when a mask test
violation occurs, or when a limit check violation in a measurement occurs.
Actions on trigger

Table of Contents

Other manuals for R&S RTO6

Related product manuals