Remote control commands
R&S
®
RTO6
2306User Manual 1801.6687.02 ─ 05
SEARch:TRIGger:MPHY:TC <SearchName>,<TC>
SEARch:TRIGger:MPHY
:TC?
<SearchName>
Specifies the TC bit pattern.
Parameters:
<TC> Numeric or string pattern, see Chapter 24.4.6, "Bit pattern
parameter", on page 1286. The string parameter accepts the bit
value X (don't care).
Parameters for setting and query:
<SearchName>
SEARch:TRIGger:MPHY:DATA:DCONdition <SearchName>,<DataOperator>
SEARch:TRIGger:MPHY:DATA:DCONdition? <SearchName>
Sets the condition for the data. You can define an exact value or a value range.
Parameters:
<DataOperator> EQUal | NEQual | LTHan | LETHan | GTHan | GETHan |
INRange | OORange
EQUal | NEQual | LTHan | LETHan | GTHan | GETHan
Equal, Not equal, Less than, Less or equal than, Greater Than,
Greater or equal than. These conditions require one data pattern
to be set with the corresponding SEARch:
TRIGger:MPHY:
DATA:DMIN command.
INRange | OORange
In range/Out of range: set the minimum and maximum value of
the range with SEARch:TRIGger:MPHY:DATA:DMIN and
SEARch:TRIGger:MPHY:DATA:DMAX.
*RST: EQUal
Parameters for setting and query:
<SearchName>
SEARch:TRIGger:MPHY:DATA:DMIN <SearchName>,<DataMin>
SEARch:TRIGger:MPHY:DA
TA:DMIN? <SearchName>
Specifies the data bit pattern, or sets the the start value of a pattern range.
Parameters:
<DataMin> Numeric or string pattern, see Chapter 24.4.6, "Bit pattern
parameter", on page 1286. The string parameter accepts the bit
value X (don't care).
Parameters for setting and query: