Remote control commands
R&S
®
RTO6
2356User Manual 1801.6687.02 ─ 05
TRIGger<m>:PCIE:ERRC:DISP <Disparity>
Enables triggering on disparity errors.
Parameters:
<Disparity> ON | OFF
*RST: OFF
TRIGger<m>:PCIE:ERRC:ECRC <ECRC>
Enables triggering on end-to-end cyclic redundancy checksum (ECRC) errors.
Parameters:
<ECRC> ON | OFF
*RST: OFF
TRIGger<m>:PCIE:ERRC:INVP <InvalidSymbol>
Enables triggering on invalid symbol errors.
Parameters:
<InvalidSymbol> ON | OFF
*RST: OFF
TRIGger<m>:PCIE:ERRC:LCRC <LCRC>
Enables triggering on link cyclic redundancy check (LCRC) errors.
Parameters:
<LCRC> ON | OFF
*RST: OFF
TRIGger<m>:PCIE:OSET:TYPE <Type>
Sets the ordered set to be triggered on.
Parameters:
<T
ype> SKP | TS1 | TS2 | FTS | EIOS | EIEOS | COMP
SKP
SKP ordered sets
TS1
Training sequence 1
TS2
T
raining sequence 2
FTS
Fast training sequence
EIOS
Electrical idle ordered set
Protocols