Remote control commands
R&S
®
RTO6
2404User Manual 1801.6687.02 ─ 05
Suffix:
<m>
.
1..4
Parameters:
<HorizontalRange> Range: 100E-12 to 100
Increment: 1E-9
*RST: 200E-9
Default unit: s
Usage: Asynchronous command
EYE<m>:HORizontal:ABSolute:POSition <HorizPosition>
Sets the place of the zero point in the diagram, in seconds. The zero point is the align-
ment point on which the slice timestamps are superimposed.
Suffix:
<m>
.
1..4
Parameters:
<HorizPosition> Range: -100 to 100
Increment: 1E-9
*RST: 0
Default unit: s
Usage: Asynchronous command
EYE<m>:HORizontal:UINTerval:RANGe <RangeUI>
Sets the number of unit intervals that are visible in the eye diagram.
Suffix:
<m>
.
1..4
Parameters:
<RangeUI> Range: 1E-6 to 100E+9
Increment: 0.01
*RST: 2
Usage: Asynchronous command
EYE<m>:HORizontal:UINTerval:POSition <PositionUI>
Sets the zero point in the diagram, in unit intervals. The zero point is the alignment
point on which the segment timestamps are superimposed.