Remote control commands
R&S
®
RTO6
2405User Manual 1801.6687.02 ─ 05
Parameters:
<PositionUI> Range: -100E+9 to 100E+9
Increment: 0.01
*RST: 0.5
Default unit: UI
Usage: Asynchronous command
EYE<m>:TIMReference:CLOCk:EDGE <EdgeType>
The clock edge sets the point in time when the state of the data signal is checked.
Suffix:
<m>
.
1..4
Parameters:
<EdgeT
ype> POSitive | NEGative | EITHer
*RST: EITHer
Usage: Asynchronous command
EYE<m>:TIMReference:CLOCk:FACTor <Factor>
Sets a clock multiplier factor
.
A
vailable if EYE<m>:TIMReference[:SOURce] is set to CLOCk.
Suffix:
<m>
.
1..4
Parameters:
<Factor> Range: 1 to 100
Increment: 1
*RST: 1
Usage: Asynchronous command
EYE<m>:TIMReference:CLOCk:FREQuency <Frequency>
Sets the clock frequency for the explicit clock signal.
Available if EYE<m>:TIMReference[:
SOURce] is set to CLOCk.
Suffix:
<m>
.
1..4
Parameters:
<Frequency> Range: 1000 to 10E+9
Increment: 10
*RST: 1E+9
Default unit: Hz
Usage: Asynchronous command