Remote control commands
R&S
®
RTO6
2406User Manual 1801.6687.02 ─ 05
EYE<m>:TIMReference:CLOCk:FREStimation
Executes an estimation of the signal for the clock frequency
.
A
vailable if EYE<m>:TIMReference[:SOURce] is set to CLOCk.
Suffix:
<m>
.
1..4
Usage: Event
EYE<m>:TIMReference:CLOCk:LEVel <Level>
Sets the vertical reference level of the timing reference source.
A
vailable if EYE<m>:TIMReference[:
SOURce] is set to CLOCk.
Suffix:
<m>
.
1..4
Parameters:
<Level> UPPer | MIDDle | LOWer
*RST: MIDDle
Usage: Asynchronous command
EYE<m>:TIMReference:CLOCk:OFEStimation
Executes an estimation of the signal for the clock offset.
Available if EYE<m>:TIMReference[:SOURce] is set to CLOCk.
Suffix:
<m>
.
1..4
Usage: Event
EYE<m>:TIMReference:CLOCk:OFFSet <Offset>
Sets a clock of
fset for the explicit clock signal.
Available if EYE<m>:TIMReference[:SOURce] is set to CLOCk.
Suffix:
<m>
.
1..4
Parameters:
<Offset> Range: -1000 to 1000
Increment: 0.01
*RST: 0
Default unit: UI
Usage: Asynchronous command