Remote control commands
R&S
®
RTO6
2407User Manual 1801.6687.02 ─ 05
EYE<m>:TIMReference:CLOCk:SAMTime <SamplingT
ime>
Sets a sampling time for the explicit clock signal, an of
fset for the clock edge in relation
to the bit start. The clock edge sets the beginning of the unit interval. The sampling
time is a number between 0 and 1. Value 0 sets the clock edge to the beginning of the
bit period; value 0.5 sets the clock edge to the middle of the bit period.
Available if EYE<m>:TIMReference[:SOURce] is set to CLOCk.
Suffix:
<m>
.
1..4
Parameters:
<SamplingTime> Range: 0 to 1
Increment: 0.01
*RST: 0.5
Default unit: UI
Usage: Asynchronous command
EYE<m>:TIMReference:CLOCk:SOURce <SourceType>
Selects the source of the clock channel, if EYE<m>:TIMReference[:
SOURce] is set
to CLOCk.
Suffix:
<m>
.
1..4
Parameters:
<SourceType> C1W1 | C2W1 | C3W1 | C4W1 | M1 | M2 | M3 | M4 | M5 | M6 |
M7 | M8 | R1 | R2 | R3 | R4 | D0 | D1 | D2 | D3 | D4 | D5 | D6 |
D7 | D8 | D9 | D10 | D11 | D12 | D13 | D14 | D15
Usage: Asynchronous command
EYE<m>:TIMReference:PROTocol:SELect <SerialBusIndex>
Selects the active bus for the eye analysis, if EYE<m>:TIMReference[:SOURce]
is
set to PROTocol.
Suffix:
<m>
.
1..4
Parameters:
<SerialBusIndex> SBUS1 | SBUS2 | SBUS3 | SBUS4
*RST: SBUS1
Usage: Asynchronous command
EYE<m>:TIMReference[:SOURce] <ReferenceSource>