Remote control commands
R&S
®
RTO6
2504User Manual 1801.6687.02 ─ 05
ADVJitter<m>:RESult:VERTical:HBINs <MaxBinNumber>
Sets the number of bins, the number of data intervals, of the histogram.
Suffix:
<m>
.
1..2
Parameters:
<MaxBinNumber> Range: 2 to 1000
Increment: 10
*RST: 200
ADVJitter<m>:RESult:VERTical:CONTinuous <State>
If enabled, the measurement scale is set automatically.
If disabled you can manually set a ADVJitter<m>:RESult:COMPonents<n>:
SCALe and a ADVJitter<m>:RESult:COMPonents<n>:OFFSet.
Suffix:
<m>
.
1..2
Parameters:
<State> ON | OFF
*RST: ON
ADVJitter<m>:RESult:VERTical:BMTYpe <MeasType>
Selects the basic measurement type to apply the vertical scale to.
Suffix:
<m>
.
1..2
Parameters:
<MeasT
ype> SBRate | SBDuration
SBRate: Symbol rate
SBDuration: Symbol duration
*RST: SBRate
ADVJitter<m>:RESult:VERTical:BHContinuous <State>
If enabled, the measurement scale of the basic jitter measurement histogram and track
is set automatically.
If disabled you can manually set a ADVJitter<m>:
RESult:COMPonents<n>:
SCALe and a ADVJitter<m>:RESult:COMPonents<n>:OFFSet.
Suffix:
<m>
.
1..2
Jitter analysis and clock data recovery