Spectrum analysis
R&S
®
RTO6
380User Manual 1801.6687.02 ─ 05
Remote command:
CALCulate:
MATH<m>:
FFT:LOGScale on page 1566
Center frequency
Defines the position of the displayed frequency range, which is (Center - Span/2) to
(Center + Span/2). The width of the range is defined using the "Frequency span" set-
ting.
Remote command:
CALCulate:MATH<m>:
FFT:CFRequency on page 1567
Frequency span
The span is specified in Hertz and defines the width of the displayed frequency range,
which is (Center - Span/2) to (Center + Span/2). The position of the span is defined
using the "Center frequency" setting.
Remote command:
CALCulate:MATH<m>:
FFT:SPAN on page 1568
Full span
Displays the full frequency span.
Remote command:
CALCulate:MATH<m>:FFT:FULLspan on page 1567
Start frequency
Defines the start frequency of the displayed frequency span.
Remote command:
CALCulate:MATH<m>:FFT:STARt on page 1566
Stop frequency
Defines the stop frequency of the displayed frequency span.
Remote command:
CALCulate:MATH<m>:FFT:STOP on page 1567
Span/RBW Coupling
Couples the frequency span to the "Resolution BW" setting.
Remote command:
CALCulate:MATH<m>:FFT:BANDwidth[:RESolution]:AUTO on page 1569