EasyManua.ls Logo

R&S RTO6 - Page 526

R&S RTO6
2677 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Protocol analysis
R&S
®
RTO6
526User Manual 1801.6687.02 ─ 05
"Data"
Trigger on a serial pattern at a defined position in the data packet.
The pattern can include several subsequent symbols (data frames).
See "Data conditions"
on page 526.
"Parity error"
Triggers on a parity error indicating a transmission error. This trigger
type is only available if a parity is configured for the UART bus.
"Break condi-
tion"
T
riggers if a start bit is not followed by a stop bit, the data line
remains at logic 0 for longer than a UART word.
"Stop error"
T
riggers if the stop bit is a logic 0.
Remote command:
TRIGger<m>:UART:TYPE on page 1733
Trigger source
Selects the transmitter or receiver line as trigger source.
Remote command:
TRIGger<m>:UART:SOURce on page 1733
Data conditions
Specify the data conditions if the trigger type is set to "Data".
Condition ← Data conditions
Selects the operator for the "Data" pattern: "Equal" or "Not equal".
Remote command:
TRIGger<m>:UART:FCONdition on page 1734
UART/RS-232/RS-422/RS-485 (option R&S RT
O6-K510)

Table of Contents

Other manuals for R&S RTO6

Related product manuals