EasyManua.ls Logo

R&S RTO6 - Page 570

R&S RTO6
2677 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Protocol analysis
R&S
®
RTO6
570User Manual 1801.6687.02 ─ 05
Error types: Item name, Enable ← Advanced
The table lists the error types you can trigger on. Select the required errors in the
"Enable" column. A
vailable are "CRC" error and "Parity" error.
Remote command:
TRIGger<m>:CMSB:ADVanced:ERENable on page 1772
TRIGger<m>:CMSB:ADVanced:ERRor<n>:ENABle on page 1772
Frame type ← Advanced
Selects the frame type for the custom bus trigger analysis.
You can define individual checking parameters for the fields listed in the "Trigger type
dependent settings".
To specify these parameters, select a field from this list and define the data and/or
index operators and values, or the bit state.
The trigger instant is the last criterion that is fulfilled.
Remote command:
TRIGger<m>:CMSB:ADVanced:FRAMe<n>:ENABle on page 1770
Field name / Condition / Enable ← Advanced
This table lists the field numbers and names in the selected frame together with a sum-
mary of the user settings of checking conditions for each field and a checkbox to
enable the checking.
Custom: Manchester / NRZ (option R&S R
T
O6-K510)

Table of Contents

Other manuals for R&S RTO6

Related product manuals