Protocol analysis
R&S
®
RTO6
644User Manual 1801.6687.02 ─ 05
"Transmission Sequence"
Triggering depends on the additional selectable "Sequence" parame-
ter:
●
"Sequence > Status": triggers at the end of the "Status" nibble.
●
"Sequence > Status+Data": triggers at the end of the last data
nibble.
Description of trigger type specific settings: "Transmission Sequence
setup"
 on page 645
"Serial Message"
Triggering on a serial message depends on the serial protocol
selected with Serial protocol > Short | Enhanced
 and the associated
setting parameters:
●
"Short" serial message
– "Sequence > Identifier": triggers at the end of the "ID".
– "Sequence > ID+Data": triggers at the end of the "ID and
Data".
Description of trigger type specific settings: "Serial Message
setup" on page 646
●
"Enhanced" 
serial message
– "Sequence > Identifier": triggers at the end of the "ID".
– "Sequence > ID+Data": triggers at the end of the "ID and
Data".
Description of trigger type specific settings: "Serial Message
setup" on page 646
SENT (option R&S RT