Protocol analysis
R&S
®
RTO6
664User Manual 1801.6687.02 ─ 05
Less than, Greater than
Less or equal, Greater or equal
In range, Out of range
Remote command:
SEARch:
TRIGger:SENT:
SICN on page 1921
Identifier min ← Serial Message setup
Defines the bit pattern of the message identifier.
In binary format, use the following characters: 1; 0; or X (do not care). To enter the pat-
tern in other formats use the bit pattern editor, see Chapter 13.1.7, "Bit pattern editor",
on page 493.
Remote command:
SEARch:TRIGger:SENT:SIMN on page 1921
Identifier max ← Serial Message setup
The second identifier pattern is required to specify a range with conditions "In range"
and "Out of range".
Remote command:
SEARch:TRIGger:SENT:SIMX on page 1921
SSM Data Condition ← Serial Message setup
Selects the operator to set a specific data pattern or a data range.
The available operators:
Equal, Not equal
Less than, Greater than
Less or equal, Greater or equal
In range, Out of range
Remote command:
SEARch:TRIGger:SENT:SDCN on page 1922
Data min ← Serial Message setup
Sets the data pattern. Enter the bytes in msb first bit order. The maximum pattern
length is 64 bit. Waveform data is compared with the pattern byte-by-byte.
Remote command:
SEARch:TRIGger:SENT:SDMN on page 1922
Data max ← Serial Message setup
The second data pattern is required to specify a range with conditions "In range" and
"Out of range".
Remote command:
SEARch:TRIGger:SENT:SDMX on page 1922
Error conditions setup
Performs the search on one or more of the following error events:
"T
ransmission sequence errors"
"Calibration Pulse Error"
Searches for calibration pulse errors.
SENT (option R&S RT
O6-K520)