EasyManua.ls Logo

R&S RTO6 - Page 1921

R&S RTO6
2677 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Remote control commands
R&S
®
RTO6
1921User Manual 1801.6687.02 ─ 05
SEARch:TRIGger:SENT:SICN <SearchName>,<SMIDOperator>
SEARch:TRIGger:SENT
:SICN?
<SearchName>
Sets the operator to set a specific data pattern or a data pattern range.
Parameters:
<SMIDOperator> EQUal | NEQual | LTHan | LETHan | GTHan | GETHan |
INRange | OORange
EQUal | NEQual | LTHan | LETHan | GTHan | GETHan
Equal, Not equal, Less than, Less or equal than, Greater than,
Greater or equal than. These conditions require one data pattern
to be set with
SEARch:TRIGger:SENT:SIMN.
INRange | OORange
In range / Out of range. To define the range set the minimum
and maximum values with SEARch:TRIGger:SENT:SIMN and
SEARch:TRIGger:SENT:SIMX.
*RST: EQUal
Parameters for setting and query:
<SearchName>
SEARch:TRIGger:SENT:SIMN <SearchName>,<SMIDPattern>
SEARch:TRIGger:SENT:SIMN?
<SearchName>
Specifies a message identifier pattern, or sets the start value of an identifier range.
Parameters:
<SMIDPattern> Numeric or string pattern, Chapter 24.4.6, "Bit pattern parame-
ter", on page 1286. The string parameter accepts the bit value X
(don't care).
Parameters for setting and query:
<SearchName>
SEARch:TRIGger:SENT:SIMX <SearchName>,<SMIDPatternTo>
SEARch:TRIGger:SENT
:SIMX? <SearchName>
Sets the end value of an identifier range for SEARch:TRIGger:SENT:SICN
INRange or OORange.
Parameters:
<SMIDPatternTo> Numeric or string pattern, Chapter 24.4.6, "Bit pattern parame-
ter", on page 1286. The string parameter accepts the bit value X
(don't care).
Parameters for setting and query:
<SearchName>
Protocols

Table of Contents

Other manuals for R&S RTO6

Related product manuals