"Byte/Bit count
(Max)"
Defines the second byte count value that is required to specify a
range with conditions
"In range"
and "Out of range".
Remote command:
TRIGger<m>:RFFE:BCOunt:CONDition on page 2064
TRIGger<m>:RFFE:BCOunt:MIN on page 2064
TRIGger<m>:RFFE:BCOunt:MAX on page 2064
Data
Defines the data trigger condition. The data setup consists of the condition and one or
two data patterns.
To define on which data frames of the sequence you want to trigger, use the Index set-
tings.
"Condition"
Sets the operator to trigger on a specific data pattern ("Equal" or "Not
equal") or a data range.
"Data (Min)"
Defines the data pattern for all operators that require one pattern.
"Data (Max)"
Defines the second data pattern that is required to specify a range
with conditions "In range" and "Out of range".
Remote command:
TRIGger<m>:RFFE:
DATA:DCON on page 2065
TRIGger<m>:RFFE:DATA:DMIN on page 2065
TRIGger<m>:RFFE:DATA:DMAX on page 2065
Index
Defines which data frames are relevant for the trigger. The setting is available for all
"Extended Register" command sequences, which can transfer more than one data
frame.
The index allows you to check for a certain bit pattern at a certain position in the byte
sequence. Furthermore, you can trigger on the occurrence of a certain bit pattern
within a data range.
To trigger on any data byte that fulfills the data pattern settings, set the index to XX.
"Condition"
Sets the operator to trigger on a specific data frame ("Equal") or a
data frame range.
"Data (Min)"
Defines the frame index of the data pattern for all operators that
require one index.
"Data (Max)"
Defines the frame index of the last data pattern that is required to
specify a range with condition
"In range"
. The instrument adjusts this
value, if you enter a data pattern that is longer than the index range.
Remote command:
TRIGger<m>:RFFE:
DATA:ICONdition on page 2065
TRIGger<m>:RFFE:DATA:IMIN on page 2066
TRIGger<m>:RFFE:DATA:IMAX on page 2066
Mask
Defines the mask trigger condition. The mask setup consists of the condition and one
or two data patterns.The setting is available for "Masked Write" command sequences.
RFFE (option R&S R