Protocol analysis
R&S
®
RTO6
803User Manual 1801.6687.02 ─ 05
"Index Max"
The second index pattern is required to specify a range with condi-
tions "In range"
.
Remote command:
TRIGger<m>:TBTO:
DATA:ICONdition on page 2129
TRIGger<m>:TBTO:DATA:IMIN on page 2130
TRIGger<m>:TBTO:DATA:IMAX on page 2130
Errors
Sets the type of error events to be triggered on.
RS-FEC error ← Errors
Triggers, if more than 22 RS-FEC errors are detected.
Remote command:
TRIGger<m>:
TBTO:ERRor:FEC on page 2131
CRC error ← Errors
Triggers on a frame that has a mismatch of the cyclic redundancy check (CRC) value
between the transmitting and receiving device.
Remote command:
TRIGger<m>:TBTO:ERRor:CRC on page 2130
Out of range error ← Errors
Triggers if values are out of range. This can be, for example a preamble error, or inva-
lid symbols in control blocks.
Remote command:
TRIGger<m>:TBTO:ERRor:OOR on page 2131
Zero terrnary symbol ← Errors
Triggers, if a zero ternary symbol pair (0-0) is detected.
Remote command:
TRIGger<m>:TBTO:ERRor:ZERO on page 2131
Ethernet 1000BASE-T1 (option R&S R