Protocol analysis
R&S
®
RTO6
826User Manual 1801.6687.02 ─ 05
See "Error condition
"
on page 828
Token
Sets the trigger to one out of four different token types:
●
"OUT"
●
"IN"
●
"SOF"
●
"SETUP"
If no additional conditions are set, the trigger instant is after the PID.
If the PID error check is selected:
●
If an error is found in the PID's complementary form, the trigger is set immediately
after the 8
th
bit of the PID
●
If no PID error is found, no trigger is set, even if other selected conditions are met.
Otherwise, the trigger instant is defined by the first occurrence of any of the specified
additional conditions.
Remote command:
TRIGger<m>:USB:TOKen on page 2160
OUT, IN or SETUP ← Token
For the trigger token types "OUT", "IN" or "SETUP", the following conditions can be
set:
You can refine the trigger condition:
●
See "PID error " on page 829
●
See "End Point" on page 829
●
See "Address" on page 829
SOF ← Token
For the trigger token type "SOF", the following conditions can be set:
You can refine the trigger condition:
●
See "PID error " on page 829
●
See "Frame number" on page 830
Data
Sets the trigger to one out of four different data types:
●
"DATA0"
●
"DATA1"
●
"DATA2"
●
"MDATA"
If no additional conditions are set, the trigger instant is after the PID.
If the PID error check is selected:
●
If an error is found in the PID's complementary form, the trigger is set immediately
after the 8
th
bit of the PID
●
If no PID error is found, no trigger is set, even if other selected conditions are met.
Otherwise, the trigger instant is defined by the first occurrence of any of the specified
additional conditions.
USB (option R&S R
T
O6-K570)