Remote control commands
R&S
®
RTO6
1652User Manual 1801.6687.02 ─ 05
Usage: Event
EXPort:WAVeform:SCOPe <Scope>
Defines the part of the waveform record that has to be stored.
Parameters:
<Scope> WFM | ZOOM | CURSor | GATE | MANual
WFM
Complete waveform
ZOOM
Data included in the zoom area if a zoom is defined for the
source waveform.
CURSor
Data between the cursor lines if a cursor measurement is
defined for the source waveform.
GA
TE
data included in the measurement gate if a gated measurement
is defined for the source waveform.
MANual
Saves the data between user-defined start and stop values to be
set with
EXPort:WAVeform:STARt and EXPort:WAVeform:
STOP.
*RST: WFM
Example: See Chapter 24.3.5.2, "Exporting waveform data to file",
on page 1267
Usage: Asynchronous command
EXPort:WAVeform:STARt <Start>
Sets the start value of the waveform section for export, if EXPort:WAVeform:SCOPe
is set to
Manual.
Parameters:
<Start> Range: -100E+24 to 100E+24
Increment: 0.01
*RST: 0.01
Default unit: s
Usage: Asynchronous command
EXPort:WAVeform:STOP <Stop>
Sets the end value of the waveform section for export, if EXPort:WAVeform:SCOPe
is set to
Manual.
Data management