EasyManua.ls Logo

R&S RTO6 - Page 1888

R&S RTO6
2677 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Remote control commands
R&S
®
RTO6
1888User Manual 1801.6687.02 ─ 05
Parameters:
<DataPattern> Numeric or string pattern, see Chapter 24.4.6, "Bit pattern
parameter", on page
1286. The string parameter accepts the bit
value X (don't care).
Parameters for setting and query:
<SearchName>
SEARch:TRIGger:FLXRay:DPOPerator
<SearchName>,<DataPosOperator>
SEARch:TRIGger:FLXRay:DPOPerator? <SearchName>
Sets the operator for the data position. Y
ou can defined an exact position, or a position
range.
Parameters:
<DataPosOperator> ANY | OFF | EQUal | GETHan | INRange | RANGe
OFF = ANY
The data position is not relevant for the search condition.
EQUal | GETHan
Equal, Greater or equal than. These conditions require one data
position to be set with SEARch:TRIGger:FLXRay:
DPOSition
INRange = RANGe
In range: Set the minimum and maximum value of the range with
SEARch:TRIGger:FLXRay:DPOSition and SEARch:
TRIGger:FLXRay:DPTO.
*RST: EQUal
Parameters for setting and query:
<SearchName>
SEARch:TRIGger:FLXRay:DPOSition <SearchName>,<DataPosition>
SEARch:TRIGger:FLXRay:DPOSition? <SearchName>
Sets the number of data bytes to be skipped after start of the payload segment.
Parameters:
<DataPosition> Range: 0 to 255
Increment: 1
*RST: 0
Parameters for setting and query:
<SearchName>
SEARch:TRIGger:FLXRay:DPTO <SearchName>,<DataPositionTo>
SEARch:TRIGger:FLXRay:DPT
O? <SearchName>
Defines the last byte of interest, if the position operator SEARch:TRIGger:FLXRay:
DPOPerator defines a range.
Protocols