Triggers
R&S
®
RTO6
231User Manual 1801.6687.02 ─ 05
"A Zone"
Triggers if the conditions of the A-event and the zone trigger are fulfil-
led. A holdof
f condition also can be set.
"A
B R
Zone"
Triggers if all conditions of A- and B-events, reset timeout or R-event
(reset) conditions are fulfilled for the defined zone.
"A OR B"
Triggers if the conditions of A-event or the conditions of the B-event
are fulfilled.
"A OR B
Zone"
Triggers if the conditions of A-event or the conditions of the B-event
are fulfilled for the defined zone. A holdof
f condition also can be set.
Remote command:
TRIGger<m>:
SEQuence:
TYPE on page 1427
After 1st event wait
Sets the time that the instrument waits after an A-trigger until it recognizes B-triggers.
Remote command:
TRIGger<m>:SEQuence:DELay on page 1428
2nd event count
Sets the number of B-trigger conditions to be fulfilled after an A-trigger. The last B-trig-
ger causes the trigger event.
The waiting time for B-triggers can be restricted with a reset condition: timeout or reset
event.
Remote command:
TRIGger<m>:SEQuence:COUNt on page 1428
Reset on event/Reset event
If enabled, the trigger sequence is restarted by the R-trigger condition if the specified
number of B-triggers does not occur before the trigger conditions are fulfilled.
The R-trigger condition is configured in the same way as the A-trigger. The instrument
disables settings that do not fit the previous settings in the sequence.
Remote command:
TRIGger<m>:SEQuence:RESet:EVENt on page 1428
Reset on timeout/Reset timeout
If timeout is enabled, the instrument waits for the "Timeout" time for the specified num-
ber of B-triggers. If no trigger occurs during that time, the sequence is restarted with
the A-trigger.
Remote command:
TRIGger<m>:SEQuence:RESet:TIMeout[:ENABle] on page 1429
TRIGger<m>:SEQuence:RESet:TIMeout:TIME on page 1429
Couple sequence thresholds
Sets the trigger levels to the values of the latest configured trigger condition. Each
channel has its own trigger level.
Sequence