Remote control commands
R&S
®
RTO6
2489User Manual 1801.6687.02 ─ 05
ADVJitter<m>:SIGNal:SAMTime <SamplingT
ime>
Sets a sampling time for the explicit clock signal, an of
fset for the clock edge in relation
to the bit start. The clock edge sets the beginning of the unit interval. The sampling
time is a number between 0 and 1. Value 0 sets the clock edge to the beginning of the
bit period; value 0.5 sets the clock edge to the middle of the bit period.
Suffix:
<m>
.
1..2
Parameters:
<SamplingTime> Range: 0 to 1
Increment: 0.01
*RST: 0.5
Default unit: UI
ADVJitter<m>:SIGNal:GATE:COUPling <ModeType>
Sets if the gate is coupled to the zoom or cursor
.
Suffix:
<m>
.
1..2
Parameters:
<ModeType> NONE | CURSor | ZOOM
*RST: NONE
ADVJitter<m>:SIGNal:GATE:CURSor <CursorType>
Sets the cursor the gate is coupled to, if
ADVJitter<m>:SIGNal:GATE:COUPling
is set to CURSor.
Suffix:
<m>
.
1..2
Parameters:
<CursorType> CURSOR1 | CURSor1 | CURSOR2 | CURSor2 | CURSOR3 |
CURSor3 | CURSOR4 | CURSor4
*RST: CURSOR1
ADVJitter<m>:SIGNal:GATE:RANGe:ABSolute:STARt <StartValue>
Sets the start time of the gate area in percent, if
ADVJitter<m>:SIGNal:GATE:
RANGe:MODE is set to ABS.
Suffix:
<m>
.
1..2
Parameters:
<StartValue> Range: -100E+24 to 100E+24
Increment: 0.01
*RST: 0
Jitter analysis and clock data recovery