436User Manual 1801.6687.02 ─ 05
Use gate
Enables the gate settings and shows the gate. Search is only performed on the defined
gate area of the source waveform.
Remote command:
SEARch:GATE[:STATe]
on page 1631
Display state
Displays the gate area in the source diagram.
Remote command:
SEARch:GATE:SHOW on page 1632
Gate description
Defines whether the gate settings are configured using absolute or relative values.
"Absolute"
The gate is defined by absolute start and stop values.
"Relative"
The gate's start and stop values are defined by a percentage of the
value range.
Remote command:
CALCulate:
MATH<m>:
FFT:GATE:MODE on page 1573
MEASurement<m>:GATE:MODE on page 1547
SEARch:GATE:MODE on page 1632
(Relative) Start
Defines the starting value for the gate.
Remote command:
CALCulate:MATH<m>:FFT:GATE:ABSolute:STARt on page 1573
CALCulate:MATH<m>:FFT:GATE:RELative:STARt on page 1574
MEASurement<m>:GATE:ABSolute:STARt on page 1547
MEASurement<m>:GATE:RELative:STARt on page 1547
SEARch:GATE:ABSolute:STARt on page 1632
SEARch:GATE:RELative:STARt on page 1633
Search gate