Remote control commands
R&S
®
RTO6
1492User Manual 1801.6687.02 ─ 05
<MeasType>
parameter value
(remote control)
Measurement
(manual opera-
tion)
Description, result Suffix
<n>
SLEFalling Slew rate on fall-
ing edge
Steepness of the falling edge: voltage difference
between the higher and lower reference level, divided by
the fall time.
49
ATAR Arithmetic for
amplitude/time
measurements
Basic arithmetic calculations with measurement values
and scalars.
50
MEASurement<m>:ENVSelect...................................................................................... 1492
MEASurement<m>:DETThreshold.................................................................................1493
MEASurement<m>:AMPT
ime:ALEVel............................................................................ 1493
MEASurement<m>:AMPTime:PFSLope......................................................................... 1493
MEASurement<m>:AMPTime:PSLope........................................................................... 1494
MEASurement<m>:AMPTime:DELay<n>:DIRection........................................................ 1494
MEASurement<m>:AMPTime:DELay<n>:ECOunt........................................................... 1495
MEASurement<m>:AMPTime:DELay<n>:LSELect.......................................................... 1495
MEASurement<m>:AMPTime:DELay<n>:SLOPe............................................................ 1495
MEASurement<m>:AMPTime:PTCount.......................................................................... 1496
MEASurement<m>:AMPTime:ESLope........................................................................... 1496
MEASurement<m>:AMPTime:CSLope........................................................................... 1496
MEASurement<m>:AMPTime:CLCK<n>:LSELect........................................................... 1497
MEASurement<m>:AMPTime:DATA<n>:LSELect............................................................1497
MEASurement<m>:AMPTime:DTOTrigger<n>:SLOPe.....................................................1497
MEASurement<m>:AMPTime:DTOTrigger<n>:LSELect................................................... 1498
MEASurement<m>:AMPTime:LCHeck<n>:VALid............................................................ 1498
MEASurement<m>:AMPTime:LCHeck<n>:LOWer:LIMit...................................................1499
MEASurement<m>:AMPTime:LCHeck<n>:UPPer:LIMit................................................... 1499
MEASurement<m>:AMPTime:LCHeck<n>:LOWer:MARGin............................................. 1499
MEASurement<m>:AMPTime:LCHeck<n>:UPPer:MARGin.............................................. 1499
MEASurement<m>:ENVSelect <EnvelopeCurve>
The command is only relevant for measurements on envelope waveforms. It selects
the envelope to be used for measurement.
Suffix:
<m>
.
1..10
See "Selection of the measurement group: MEASurement<m>"
on page
1479.
Parameters:
<EnvelopeCurve>
MIN | MAX | BOTH
MIN: measures on the lower envelope
MAX: measures on the upper envelope
BOTH: the envelope is ignored and the waveform measured as
usual
*RST: BOTH
Usage: Asynchronous command
Automatic measurements