Remote control commands
R&S
®
RTO6
1654User Manual 1801.6687.02 ─ 05
EXPort:WAVeform:DLOGging <DataLogging>
The command enables the export of subsequent acquisitions of the selected wave-
forms. The waveforms are taken from a running Nx Single acquisition (data logging,
history is disabled), or from the history (multiple waveforms, history is enabled).
If the history is disabled (CHANnel<m>[:
WAVeform<n>]:
HISTory[:STATe]) and
data logging is enabled, a specified number of waveforms is transferred to file directly
during RUN Nx SINGLE acquisition. Enabling data logging stops a running acquisition.
Set the number of acquisitions to be acquired and stored with ACQuire:COUNt and
start export using RUNSingle.
If the history is enabled, the subsequent waveforms are taken from the history. Specify
the range with CHANnel<m>[:WAVeform<n>]:HISTory:STARt and
CHANnel<m>[:WAVeform<n>]:HISTory:STOP. Then play the history with
CHANnel<m>[:WAVeform<n>]:HISTory:PLAY.
The commands EXPort:WAVeform:SAVE, CHANnel<m>[:WAVeform<n>]:DATA[:
VALues]? and CHANnel<m>[:WAVeform<n>]:DATA:HEADer? are not available if
data logging is enabled. The RUNContinous command disables data logging.
If data logging is off, and the history is enabled, one waveform out of the history is writ-
ten to file. Specify the waveform using CHANnel<m>[:WAVeform<n>]:HISTory:
CURRent and save it using EXPort:WAVeform:SAVE.
Parameters:
<DataLogging> ON | OFF
*RST: OFF
Example: See:
"Exporting multiple running acquisitions of a single waveform to
XML file" on page
1270
"Exporting multiple acquisition of the history to XML file"
on page
1271
"Exporting a single acquisition of the history to BIN file"
on page 1270
Usage: Asynchronous command
EXPort:WAVeform:TIMestamps <UseTimestamps>
Exports the relative timestamps of all history waveforms to the waveform data file. The
time is written at the beginning of each waveform record.
Parameters:
<UseT
imestamps> ON | OFF
*RST: OFF
Usage: Asynchronous command
Data management