Protocol analysis
R&S
®
RTO6
836User Manual 1801.6687.02 ─ 05
You can refine the search condition:
●
See "Address check" on page
836
●
See
"End Point check" on page 836
Remote command:
SEARch:TRIGger:USB:SPEC on page 2178
Error condition
Searches for the following error conditions:
"PID Error"
Searches for packet identifier errors.
Remote command:
SEARch:TRIGger:USB:
PIDerror on page 2175
"CRC5 Error"
Searches for any CRC5 error event.
Remote command:
SEARch:TRIGger:USB:
CRC5error on page 2170
"CRC16 Error"
Searches for any CRC16 error event.
Remote command:
SEARch:TRIGger:USB:
CRC16error on page 2170
"Bitstuff Error"
Searches for bitstuff errors, thus an erroneous or missing bit stuffing
sequence (see USB standard).
Remote command:
SEARch:TRIGger:USB:BITSterror on page 2170
"Glitch Error"
Searches for glitch errors (errors in the bit period, see USB standard
for the definition of glitch).
Remote command:
SEARch:TRIGger:USB:
GLITcherror on page 2174
Address check
Enables the search for an address that meets specific conditions.
"Condition"
Defining a specific address or an address range requires to set the
operator to one of the following: equal, not equal, less than, less than
or equal, greater than, greater than or equal, in range, out of range.
"Min"
Specifies an address, or sets the start value of an address range.
"Max"
Sets the end value of an address range if "Condition" (SEARch:
TRIGger:USB:ACONdition) is set to
INRange or OORange.
Remote command:
SEARch:TRIGger:USB:
WADD on page 2180
SEARch:TRIGger:USB:ACONdition on page 2169
SEARch:TRIGger:USB:AMIN on page 2169
SEARch:TRIGger:USB:AMAX on page 2170
End Point check
Enables the search for an endpoint that meets specific conditions.
USB (option R&S R
T
O6-K570)