Protocol analysis
R&S
®
RTO6
916User Manual 1801.6687.02 ─ 05
"Data Max"
The second data pattern is required to specify a range with conditions
"In range" and
"Out of range".
Remote command:
SEARch:TRIGger:MPHY:
DATA:DCONdition on page 2306
SEARch:TRIGger:MPHY:DATA:DMIN on page 2306
SEARch:TRIGger:MPHY:DATA:DMAX on page 2307
Index
Index sets the range within this series of data that is considered for the search. The
index setup consists of the condition and one or two data index values.
"Condition"
Defines the operator to set a specific index, e.g. "Equal" or a range.
"Index Min/Index"
Defines the bit pattern of the index pattern.
In binary format, use the following characters: 1; 0; or X (do not care).
To enter the pattern in other formats use the bit pattern editor
, see
Chapter 13.1.7, "Bit pattern editor", on page 493.
"Index Max"
The second index pattern is required to specify a range with condi-
tions "In range".
Remote command:
SEARch:TRIGger:MPHY:
DATA:ICONdition on page 2307
SEARch:TRIGger:MPHY:DATA:IMIN on page 2307
SEARch:TRIGger:MPHY:DATA:IMAX on page 2308
FSN
Sets the specified Frame Sequence Number (FSN) to be searched for. The FSN setup
consists of the condition and one or two FSN patterns.
"Condition"
Defines the operator to set a specific FSN, e.g. "Equal" or "Not Equal"
or a range.
"Data Min/Data"
Defines the bit pattern of the FSN pattern.
In binary format, use the following characters: 1; 0; or X (do not care).
To enter the pattern in other formats use the bit pattern editor, see
Chapter 13.1.7, "Bit pattern editor", on page 493.
"Data Max"
The second FSN pattern is required to specify a range with conditions
"In range" and "Out of range".
Remote command:
SEARch:TRIGger:MPHY:
FSNumber:CONDition on page 2308
SEARch:TRIGger:MPHY:FSNumber:MIN on page 2308
SEARch:TRIGger:MPHY:FSNumber:MAX on page 2309
CRC
Sets the specified Cyclic Redundancy Check (CRC) to be searched for. The CRC
setup consists of the condition and one or two CRC patterns.
CRC is an error detecting code to detect accidental changes to raw data.
M-PHY and USB SSIC (option R&S RT
O6-K580 and K570)